Place clean and dry products (handpieces, attachments, sterilization guides) into designated locations in the sterilization container. Fasten sterilization case lid.
Sterilization Container Dimensions: 425 mm (Length) x 305 mm (Width) x 120 mm (Height)
WARNING: Instruments are manufactured as non-sterile. They need to be sterilized before using.
CAUTION: ACF Medical recommends you to autoclave your tools and accessories in ACF sterilization container. Every instrument should be sterilized independent of each other.
Sterilization Recommendations
In case of any urgent need;
Parameters for Sterilizing
WARNING: Instruments are manufactured as non-sterile. They need to be sterilized before using.
CAUTION: The following value must not be exceeded for sterilizing: 134° C. Higher values can damage the sterilized products.
CAUTION: Do not use the handpiece just after autoclave process till temperature of trigger mechanism to cool down and reach to room temperature. If you use it before cooling; handpiece will be damaged.
CAUTION: Products such as coolant should not be used to speed up the cooling of the instruments.
CAUTION: Where there is a concern about TSE/vCJD contamination, the World Health Organization recommends processing through a pre-vacuum steam sterilization cycle for 18 minutes at 134 °C (273 °F). (WHO/CDS/CSR/ APH 2000.3, “WHO Infection Control Guidelines for TSE,” March 1999).