BATTERIES ARE NON-STERILE AND NOT AUTOCLAVABLE
DO NOT LEAVE THE BATTERIES IN THE HANDPIECE
PLEASE CHARGE BEFORE AND AFTER EACH SURGERY
PLEASE CHARGE BEFORE STORING
The batteries have lithium-ion technology. A02 520 Battery Li-Ion is rechargeable up to 800 and 1000 times depending on utilization. After 500 times charging, as with all lithium-Ion batteries, the battery life will be reduced. (Such as our mobile phones) A normal battery is designed to retain up to 80% of its original capacity at 500 complete charge cycles when operating under normal conditions.
If the one battery rating is reduced under %20 battery rating, this battery can not be charged anymore. Because this is the minimum limit for recharging of the battery. If the battery volt is under 10.8 V, this is meaning the battery died.
How the customers can reach %20 battery rate?
1 – If they use the one battery in 3 arthroplasty or trauma operations in one time without charging,
2 – If they forget or leave the battery in the handpiece,
3 – If they didn’t charge the battery after a long operation before storing,
4 – If they didn’t charge the batteries for a long time while storing.
5 – If they put the battery to the autoclave machine.
Place A02 460 Charger Li-Ion on the appropriate / flat / dry ground sufficient distance according to the operation type, insert the power cable as appropriate 220 Volts into the slot on the back of the unit.
WARNING: A02 460 Charger Li-Ion is non sterile and not autoclavable.
WARNING: In countries where the mains voltage is not 220-240 V, use a voltage converter or please inform the manufacturer before purchase the products to change the charger voltage to your country mains voltage.
CAUTION: Make sure the position On Off Switch is “O” ( Off ) before inserting plug of charger to the socket.
The battery is partially charged. To ensure full capacity, you need to completely charge the battery with the battery charger before using your power tool for the first time. The lithiumion battery can be charged at any time without reducing its service life. Interrupting charging procedure does not damage the battery.
After turning on the charger, insert the A02 520 Battery Li-Ion to the charging bay.
“READY” LED lights up as green when the battery is placed and the percentage of the battery rates appear on the LCD screen.
When the batteries are full, “FUL” led light will begin flashing and “FUL” will appear on the screen. A02 520 Battery Li-Ion is ready to insert to the handpiece.
There are 3 error codes for different battery malfunctions to see on screen of charger.
ER1 – The battery fuse is blown or one or more battery cells are dead.
ER2 – The batteries did not reach full charge during the ideal charging time.
ER3 – The batteries are under the appropriate charging voltage and cannot be charged.
CAUTION: When “ERROR” led light is flashed, please get in contact with us.
WARNING: Return the device to the manufacturer if there is visible dirtiness on it.
CAUTION: Normally, one fully charged battery has sufficient capacity for an entire operation. As a precaution, a second battery should be kept ready. Charge only with “A02 460 – Charger Li-Ion”. The batteries should always be charged before use. Place the battery into the charger immediately after surgery. Keep the charger and the batteries clean and in a cool and dry place.
Recycling: Recycle the batteries according to national and local regulations.
CAUTION: If the non-sterile battery contacts the outside of the handpiece, the handpiece must be cleaned and resterilized before being used.
Circulating Person (Non-Sterile Person)
• Insert the battery through guide into the handpiece.
• The circulating person must not touch the outside of the handpiece.
CAUTION: Inserting the batteries very quickly can cause system protection and the power tool may not work. Please wait a few seconds and reinsert the battery.
CAUTION: Be sure the A02 520 Battery Li-Ion is full before inserting to the handpiece. If the battery rate is under %20, the handpiece won’t work. The batteries must be definitely charged before and after each operation.
CAUTION: Remove the battery from the handpiece and charge it after each operation. Do not leave the batteries in the handpieces. If you leave it, after a while battery life will start to be shorten and battery will malfunction.
CAUTION: Do not pull the trigger while inserting the battery. Ensure that your finger does not touch the trigger.
CAUTION: Avoid all contact with the outside of the handpiece in order not to contaminate it. Should the non-sterile battery or the circulating person’s hand come into contact with the outside of the handpiece, it must be cleaned and resterilized before being used.
Scrubbed Person (Sterile Person)
• Close the lid and turn the latch to “closed” situation. The power tool is ready for operation.
CAUTION: Do not contact the non-sterile battery in order to avoid contamination. Should the scrubbed person come into contact with the non-sterile battery, they must be scrubbed again. Should the handpiece be contaminated it must be cleaned and resterilized before being used.
CAUTION: Before running the tool, please wait a few seconds (max 10 seconds) for optimizing (trigger, motor and voltage control etc.) after insterting the battery to the handpiece. If you don’t, you must reinsert the battery.